Author :
Rameshpetchi Rajendran, Devikasubramaniyan*, Ramya Sri Sura, Ch.Brammam, G.Shyamala, J.Swarna, J.Akhil, K.Sunajani
Volume :
5
Issue :
04
Abstract :
A rapid and precise Reverse Phase High Performance Liquid Chromatographic method has been developed for the
validated of Granisetron, in its pure form as well as in tablet dosage form. Chromatography was carried out on a
Symmetry C18 (4.6 x 250mm, 5µm) column using a mixture of Acetonitrile and Water (50:50% v/v) as the mobile
phase at a flow rate of 0.8ml/min, the detection was carried out at 285nm. The retention time of the Granisetron was
3.0 ±0.02min. The method produce linear responses in the concentration range of 20-100µg/ml of Granisetron. The
method precision for the determination of assay was below 2.0%RSD. The method is useful in the quality control of
bulk and pharmaceutical formulations.
Keywords: Granisetron, RP-HPLC, validation.
Keyword :
Granisetron, RP-HPLC, validation.