To evaluate the functional efficacy of various light curing units in dental clinics across Chhattisgarh state


Article PDF :

Veiw Full Text PDF

Article type :

Original Article

Author :

Yogesh Sahu, Vaibhav Kridutt, Aditi Jain, Pooja Deshmukh, Deepti Jain, Shruti Sharma

Volume :

2

Issue :

1

Abstract :

Introduction: A clinical survey to evaluate functional efficacy of various light curing units in dental offices across Chhattisgarh state, India. Aims: The purpose of this study was to evaluate the output intensity & the factors affecting functional efficacy of various light curing units used in private dental office. Methods and Material: The output intensity of 110 light curing units in various dental clinics across Chhattisgarh state was examined with the help of Bluephase® radiometer. Various factors were also selected and recorded that can affect the functional efficacy of light cure units. The average output intensity was divided into two categories for QTH (300mW/cm2) and for LED ( 600 mW/cm2). Results: Among the QTH light cure units, 32 out of 41 (78%) were found out to be satisfactory while 44 LED light cure units out of 69 (63.7%) were satisfactory based on the criteria used for the study (p- value= 0.12). Conclusions: observation found that there is general lack of regular maintenance of these light curing units affects the functional efficacy of these units. Key words: LED, QTH, Radiometer, Light curing units, Light intensity